	Infrastructure
	
	
	 
	Annealing station 
	
	An oil free vacuum station based on turbo molecular pump and diaphragm pump
	is set up in the GDA laboratory for annealing and evacuating the cryostats
	of HPGe detectors. This system is protected against power failures with
	solenoid activated valve. This annealing system has become the work station
	for HPGe detector maintenance activities.  | 
	
  
	  |  
	
	 
	
	
	 
	LN2 autofilling system 
	
	An automatic liquid nitrogen filling system has been designed and made in-house,
	with a dedicated controller which controls a set of electro-pneumatic
	valves/sensors to fill the 12 HPGe detectors in proper sequence and in failsafe
	mode. The temperature readout from the sensors (platinum resistor - PT100 whose
	resistance varies linearly with temperature) is communicated to the PC through
	the ADC in the controller. The controller also has the relays to operate the
	24 V dc solenoids of electro-pneumatic valves.  | 
	
 
  
	  |  
	
	 
	
	
	
	The filling system consists of dedicated manifolds for liquid nitrogen and
	compressed air on both sides of the beamline each catering to 6 detectors. The
	detectors are periodically and automatically filled once a day. Fill log is
	maintained in a logfile for monitoring purposes. Filling is done from a 200 l
	capacity LN2 dewar (at a pressure of 22 psig) which itself gets
	filled from a 5000 l dewar
	kept outside the Beam Hall, once in 2 days. The process is controlled by a GUI
	code working in linux environment. The configuration of the number of
	detectors, sensors, valves (enabled/disabled) are kept in the file
	'lnfill.conf'. 
	
	 
	
	 |  
	